In electrochemistry, the Randles–Ševčík equation describes the effect of scan rate on the peak current for a cyclic voltammetry experiment. For simple redox events where the reaction is electrochemically reversible, and the products and reactants are both soluble, such as the ferrocene/ferrocenium couple, depends not only on the concentration and diffusional properties of the electroactive species but also on scan rate.
Or if the solution is at 25 °C:
- = current maximum in amps
- = number of electrons transferred in the redox event
- = electrode area in cm2
- = Faraday constant in C mol−1
- = diffusion coefficient in cm2/s
- = concentration in mol/cm3
- = scan rate in V/s
- = Gas constant in J K−1 mol−1
- = temperature in K
- The constant with a value of 2.69×105 has units of C mol−1 V−1/2
Derivation
This equation is derived using the following governing equations and initial/boundary conditions:- = distance from a planar electrode in cm
- = time in seconds
- = the potential of the electrode in volts
- = the initial potential of the electrode in volts
- = the formal potential for the reaction between the oxidized and reduced species
Uses
Using the relationships defined by this equation, the diffusion coefficient of the electroactive species can be determined. Linear plots of ip vs. ν1/2 and peak potentials that are not dependent on ν provide evidence for an electrochemically reversible redox process. For species where the diffusion coefficient is known, the slope of the plot of ip vs. ν1/2 provides information into the stoichiometry of the redox process, the concentration of the analyte, the area of the electrode, etc.A more general investigation method is the plot of the peak currents as function of the scan rate on a logarithmically scaled x-axis. Deviations become easily detectable and the more general fit formula
can be used.
In this equation is the current at zero scan rate at the equilibrium potential. In the electrochemical lab experiment may be small but can nowadays easily be monitored with a modern equipment. For example corrosion processes may lead to a not vanishing but still detectable . When and x is close to 0.5 a reaction mechanism according to Randles Sevcik can be assigned.
An example for this kind of reaction mechanism is the redox reaction of species as an analyte in a highly concentrated background solution on graphite electrode.
A more detailed plot with all fit parameters can be seen