Energy-dispersive X-ray spectroscopy
Energy-dispersive X-ray spectroscopy, sometimes called energy dispersive X-ray analysis or energy dispersive X-ray microanalysis, is an analytical technique used for the elemental analysis or chemical characterization of a sample. It relies on an interaction of some source of X-ray excitation and a sample. Its characterization capabilities are due in large part to the fundamental principle that each element has a unique atomic structure allowing a unique set of peaks on its electromagnetic emission spectrum. The peak positions are predicted by the Moseley's law with accuracy much better than experimental resolution of a typical EDX instrument.
To stimulate the emission of characteristic X-rays from a specimen a beam of electrons or X-ray is focused into the sample being studied. At rest, an atom within the sample contains ground state electrons in discrete energy levels or electron shells bound to the nucleus. The incident beam may excite an electron in an inner shell, ejecting it from the shell while creating an electron hole where the electron was. An electron from an outer, higher-energy shell then fills the hole, and the difference in energy between the higher-energy shell and the lower energy shell may be released in the form of an X-ray. The number and energy of the X-rays emitted from a specimen can be measured by an energy-dispersive spectrometer. As the energies of the X-rays are characteristic of the difference in energy between the two shells and of the atomic structure of the emitting element, EDS allows the elemental composition of the specimen to be measured.
Equipment
Four primary components of the EDS setup are- the excitation source
- the X-ray detector
- the pulse processor
- the analyzer.
Hazards and Safety
- High Voltage: SEM-EDX operates at high voltages, which can pose a risk of electric shock.
- X-ray Radiation: While SEM-EDX does not use as high a voltage as some X-ray techniques, it still produces X-rays that can be harmful with prolonged exposure. Proper shielding and safety measures are necessary.
- Sample Preparation: Handling and preparation of samples can involve hazardous chemicals or materials. Proper personal protective equipment should be used.
- Vacuum System: The vacuum system used in SEM-EDX can implode if not properly maintained, leading to potential hazards.
- Cryogenic Hazards: Some samples may require cryogenic techniques for analysis, which can pose risks of cold burns or asphyxiation if not appropriately handled.
- Mechanical Hazards: If used incorrectly, moving parts in the SEM can cause injury.
- Fire and Explosion Risks: Some samples, particularly those involving flammable materials, can pose fire or explosion risks under vacuum conditions.
- Ergonomic Risks: Prolonged use of SEM-EDX can lead to ergonomic hazards if the workstation is not correctly set up for the user's comfort and safety.
Technological variants
X-ray photoelectron spectroscopy is another close relative of EDS, utilizing ejected electrons in a manner similar to that of AES. Information on the quantity and kinetic energy of ejected electrons is used to determine the binding energy of these now-liberated electrons, which is element-specific and allows chemical characterization of a sample.
EDS is often contrasted with its spectroscopic counterpart, wavelength dispersive X-ray spectroscopy. WDS differs from EDS in that it uses the diffraction of X-rays on special crystals to separate its raw data into spectral components. WDS has a much finer spectral resolution than EDS. WDS also avoids the problems associated with artifacts in EDS.
A high-energy beam of charged particles such as electrons or protons can be used to excite a sample rather than X-rays. This is called particle-induced X-ray emission or PIXE.
Accuracy
EDS can be used to determine which chemical elements are present in a sample, and can be used to estimate their relative abundance. EDS also helps to measure multi-layer coating thickness of metallic coatings and analysis of various alloys. The accuracy of this quantitative analysis of sample composition is affected by various factors. Many elements will have overlapping X-ray emission peaks. The accuracy of the measured composition is also affected by the nature of the sample. X-rays are generated by any atom in the sample that is sufficiently excited by the incoming beam. These X-rays are emitted in all directions, and so they may not all escape the sample. The likelihood of an X-ray escaping the specimen, and thus being available to detect and measure, depends on the energy of the X-ray and the composition, amount, and density of material it has to pass through to reach the detector. Because of this X-ray absorption effect and similar effects, accurate estimation of the sample composition from the measured X-ray emission spectrum requires the application of quantitative correction procedures, which are sometimes referred to as matrix corrections.Emerging technology
There is a trend towards a newer EDS detector, called the silicon drift detector. The SDD consists of a high-resistivity silicon chip where electrons are driven to a small collecting anode. The advantage lies in the extremely low capacitance of this anode, thereby utilizing shorter processing times and allowing very high throughput. Benefits of the SDD include:- High count rates and processing,
- Better resolution than traditional Si detectors at high count rates,
- Lower dead time,
- Faster analytical capabilities and more precise X-ray maps or particle data collected in seconds,
- Ability to be stored and operated at relatively high temperatures, eliminating the need for liquid nitrogen cooling.
- Minimizing SEM beam current allowing for optimization of imaging under analytical conditions,
- Reduced sample damage and
- Smaller beam interaction and improved spatial resolution for high speed maps.
In recent years, a different type of EDS detector, based upon a superconducting microcalorimeter, has also become commercially available. This new technology combines the simultaneous detection capabilities of EDS with the high spectral resolution of WDS. The EDS microcalorimeter consists of two components: an absorber, and a superconducting transition-edge sensor thermometer. The former absorbs X-rays emitted from the sample and converts this energy into heat; the latter measures the subsequent change in temperature due to the influx of heat. The EDS microcalorimeter has historically suffered from a number of drawbacks, including low count rates and small detector areas. The count rate is hampered by its reliance on the time constant of the calorimeter's electrical circuit. The detector area must be small in order to keep the heat capacity small and maximize thermal sensitivity. However, the count rate and detector area have been improved by the implementation of arrays of hundreds of superconducting EDS microcalorimeters, and the importance of this technology is growing.