Monitor unit


A monitor unit is a measure of machine output from a clinical accelerator for radiation therapy such as a linear accelerator or an orthovoltage unit. Monitor units are measured by monitor chambers, which are ionization chambers that measure the dose delivered by a beam and are built into the treatment head of radiotherapy linear accelerators.

Calibration and dose quantities

Linear accelerators are calibrated to give a particular absorbed dose under particular conditions, although the definition and measurement configuration may vary among medical clinics.
The most common definitions are:
  1. The monitor chamber reads 100 MU when an absorbed dose of 1 gray is delivered to a point at the depth of maximum dose in a water-equivalent phantom whose surface is at the isocenter of the machine with a field size at the surface of 10 cm × 10 cm.
  2. The monitor chamber reads 100 MU when an absorbed dose of is delivered to a point at a given depth in the phantom with the surface of the phantom positioned so that the specified point is at the isocentre of the machine and the field size is 10 cm × 10 cm at the isocentre.
Some linear accelerators are calibrated using source-to-axis distance instead of source-to-surface distance, and calibration may vary depending on hospital custom.
Early radiotherapy was performed using "constant SSD" treatments, and so the definition of monitor unit was adopted to reflect this calibration geometry.
Modern radiotherapy is performed using isocentric treatment plans, so newer definitions of the monitor unit are based on geometry at the isocenter based on the source-to-axis distance.

Secondary monitor unit calculations

Nearly 60% of the reported errors involved a lack of an appropriate independent secondary check of the treatment plan or dose calculation
With the development and technological advances, radiotherapy requires that high doses of radiation are delivered to the tumor with increasing precision. According to the recommendations of the International Commission on Radiation Units and Measurements in Publication 24
, the delivered dose should not deviate by more than ± 5% of the prescribed dose. More recently, the new ICRU recommendations in Publication 62
Commercially available computerized treatment planning systems are often used in radiotherapy services to perform monitoring unit calculations to deliver the prescribed dose to the patient. As only a part of the total dose uncertainty originates from the calculation process in treatment planning, the tolerance for accuracy of planning systems has to be smaller.
Publications on quality assurance in radiotherapy have recommended routine checks of MU calculations through independent manual calculation. This type of verification can also increase confidence in the accuracy of the algorithm and in the data integrity of the beams used, in addition to providing an indication of the limitations of the application of conventional dose calculation algorithms used by planning systems. Table 1 lists MU calculation software manufacturers.
ManufacturerSoftwareAlgorithmSupportedWebsite
LAP LaserRadCalcModified ClarksonIMRT
VMAT
TomoTherapy
CyberKnife
Halcyon
https://www.lap-laser.com/
Varian Medical SystemsMobiusCollapsed Cone Convolution/SuperpositionIMRT
VMAT
TomoTherapy
CyberKnife
Halcyon
https://www.varian.com/
Standard ImagingIMSureThree Source ModelIMRT
VMAT
https://www.standardimaging.com/
PTW FreiburgDiamondModified ClarksonIMRT
VMAT
https://www.ptwdosimetry.com/en/
Sun NuclearDoseCHECKCollapsed Cone Convolution/SuperpositionIMRT
VMAT
TomoTherapy
Halcyon
https://www.sunnuclear.com/
RT Medical SystemsRT ConnectModified Clarkson3D, 2D
IMRT
VMAT
TomoTherapy
CyberKnife
Halcyon
http://rtmedical.com.br/
Math Resolutions LLCDosimetryCheckModified ClarksonIMRT
VMAT
http://www.mathresolutions.com/
MUCheckOncology Data SystemsModified ClarksonIMRT
VMAT
TomoTherapy
CyberKnife
https://mucheck.com/odshome/