Inferior vena cava filter
An inferior vena cava filter is a medical device made of metal that is implanted by vascular surgeons or interventional radiologists into the inferior vena cava to prevent a life-threatening pulmonary embolism or venous thromboembolism.
The filter is designed to trap a blood clot and prevent its travel to the lung where it would form a pulmonary embolism. Their effectiveness and safety profile is well established, and they may be used when anticoagulant treatment is not sufficient.
Results from the PREPIC study and other studies which have shown many long-term complications of IVC filters led to the introduction of retrievable IVC filters. The first retrievable IVC filters were approved by FDA in 2003 and 2004.
In 2012, the American College of Chest Physicians recommended IVC filters for those with contraindications to anticoagulation who either have acute PE or acute proximal deep vein thrombosis.
History
The first IVC filter was created by Kazi Mobin-Uddin who published his findings in 1969 in the New England Journal of Medicine. The Mobin-Uddin filter was later replaced by the Greenfield filter developed by Lazar Greenfield which had a lower rate of filter related complications.Medical uses
While the ability to retrieve a filter does exist for many models, it cannot be guaranteed that all cases of filter placement will allow for, or be indicated for retrieval. Thus, the requirements and indications for permanent placement of filters is used to decide on when to use both permanent and temporary IVC filters.Long-term risk factors must be considered as well, to include life expectancy of more than six months following insertion, and the ability of the patient to comply with anticoagulation therapy. The decision to use a filter that is temporary vs permanent basically is tied to the expected duration of time that protection is needed to prevent pulmonary emboli from passing to the heart and lungs. One such guideline is outlined below:
- Contraindications to anticoagulation; e.g. a patient with DVT or PE who has another condition that puts them at risk of bleeding, such as a recent bleed into the brain, or a patient about to undergo major surgery
- Short-term risk of PE/Short-term contraindication of anticoagulation: Usually merits a retrievable filter
- Uncertain risk of PE and/or lack of control for anticoagulation: Usually results in permanent filters for long term management
- Long-term risk of PE/recurrent PE/recurrent DVT: permanent filter
Indication categories
Absolute indications
These are patients that should strongly consider having IVC filter placement, as they are at greatest risk of pulmonary embolus.- Proven VTE: Venous thromboembolism and contraindication or complication due to anticoagulation therapy
- Recurrent VTE: Despite adequate anticoagulation therapy
Relative indications
- Proven VTE: High risk of contraindication or complication to arise during anticoagulation therapy
- Large, free-floating proximal DVTs
- Poor compliance: INR levels are not stable, not taking medicine as directed
- Thrombolysis: Iliocaval DVT's, which are emboli in the Illiac region
Prophylactic indications
- No VTE: Anticoagulation therapy is not possible
- Transient risk of VTE: Trauma, surgical procedures or medical conditions
Other indications
- Bariatric patients: Undergoing surgery for weight control, only if BMI greater than 55, previous history of DVT/PE, hypercoagulable state, chronic venous insufficiency, truncal obesity or contraindication to anticoagulation therapy.
Indications for removing IVC filter
- No need for permanent filter: Remove temporary
- Low risk of significant PE: Continued anticoagulation is working, remove temporary
- No expected near-term high-risk PE: Continuation of anticoagulation therapy, remove temporary
- Life expectancy of more than six months: Remove temporary
- Ability to retrieve the filter: No complications, no tear probability, no trauma probability; if so, remove temporary
Anticoagulation
In those with initial acute proximal DVT or acute PE who had IVC filters placed instead of anticoagulation, and who have their risk of bleeding resolve, the American College of Chest Physicians suggested, in 2012, that they receive a standard course of anticoagulation. While IVC filters are associated with a long term risk of DVT, they are not, alone, reason enough to maintain extended anticoagulation.Side effects
The main function of a vena cava filter is to prevent death from massive pulmonary emboli. Long-term clinical follow-up studies have shown that this is accomplished in 96% of cases having a standard stainless-steel Greenfield filter.FDA communication
In August, 2010, the FDA released an Initial Communication on the Risk and Adverse events associated with Long Term use of an inferior vena cava filter. Over a period of about 5 years, they identify 921 events. While not the majority of cases, that number still represents a statistical significance of the use of long-term IVCs.Of these IVC filter side effects, 328 involved device migration, 146 involved embolizations after detachment of device components, 70 involved perforation of the IVC, and 56 involved filter fracture. Much of the medical community believes that this large number of adverse events is related to the heart filter remaining in place for longer than necessary.
Common issues relating to failure, to include death include:
- Device-associated morbidity
- Device migration
- Filter embolization
- Filter fracture
- Insertion-site thrombosis
- Perforation of the vena cava
- Recurrent DVT
- Recurrent PE
- Thrombotic complications
- Vena cava thrombosis
Less common adverse reactions
- 9% delayed filter penetration of the IVC greater than 3 mm
- Parts of filter broke off in 2%
- Caval filter migrated to heart or pulmonary artery
- Filter fracture: Broken struts migrate to retroperitoneum, requiring exploration. Also struts can migrate to the heart and can cause pericardial effusion and tamponade.
- Perforation into the duodenum: Resulting in severe diarrhea and weight loss.
- Arterial hemorrhage: Requires surgery to avoid death.
- Misplaced Filter: Causes pericardiac tamponade requiring surgery under cardiopulmonary bypass.
- Filter Migration to Right Atrium: Resulting in acute myocardial infarction.
- Filter Lodged in Heart: Causes life-threatening arrhythmia and often requires pacemakers to resolve condition.
- Heart Migration: Requires surgery to remove IVC from heart.
- Sudden Death: Caused by the filter migrating to an active region in the heart.
- Migration of Filter to Chest: Requires surgery.
- Perforation of filter strut into small-bowel: Requires surgery to repair perforation.
MRI
While most IVC filters are made of non-ferromagnetic materials, there are a few types that are weakly ferromagnetic. Accordingly, IVC filters fall under the MRI Safe and MRI Conditional categories depending mostly on type of material used during construction. Rarely will one find an MRI Not Safe IVC filter, as most of the steel, and other ferromagnetic material devices have been discontinued via the FDA.IVC filters are attached to the vena cava via hooks on their ends. Some are compression springs, which compress outward onto the side wall of the vena cava; however, they still have small hooks that retain their location. These hooks aid in the anchoring and healing process, as they allow the tissues to 'ingrow' around them, securing the IVC in place. It is unlikely, then, after 4 to 6 weeks of healing, that an MRI of 1.5 tesla, up to 3 tesla, will cause any level of dislodging to occur to the IVC filter.
Studies of MR examination of both animals and humans, with implanted IVC filters, have not reported complications or symptomatic filter displacement.
Several animal studies have even used "real-time" MR for the placement of IVC filters to check for rotation, sheering, and other artifacts.
As part of the 'routing' survey for MRI studies, patients who have IVC filters will often need verification from the doctor, or medical records, to state that the IVC is safe for the MRI. Most patients with weakly or non-safe ferromagnetic implants will be given a card, which they keep on their person at all time, that can help isolate if it is safe to do an MRI.
For patients who have been denied MRI scans for safety reasons, doctors usually recommend the CT scan with contrast as an alternative.