MiroSurge
MiroSurge is a presently prototypic robotic system designed mainly for research in minimally invasive telesurgery. In the described configuration, the system is designed according to the master slave principle and enables the operator to remotely control minimally invasive surgical instruments including force/torque feedback. The scenario is developed at the Institute of Robotics and Mechatronics within the German Aerospace Center.
The system consists of
- Three to five MIRO robot arms at the operation table,
- Minimally invasive instruments,
- a HD Stereo endoscope,
- surgical workstation with two force/torque reflecting input devices and stereo vision, and
- a planning suite for the robotic setup.
MIRO robot arm
MIRO is a lightweight robot with an optimized design for surgical applications.- Payload approximately 3 kg
- 7 degrees of freedom (DoF)
- Joint-sided torque sensors
- 3 kHz Cartesian control cycle
MICA minimally invasive instrument
MICA is a robotic instrument for minimally invasive surgery.- intra corporal universal joint
- scissors, gripper, or a Maryland as a functional tip
- Joint-sided 7DoF force/torque sensor for force feedback
- 3 kHz position control cycle
Surgical workstation
For manipulation, the surgeon sits in front of a user console, remotely located from the operation table. The console comprises- stereo vision of the site of operation on a HD-display and
- two commercially available, fully actuated input devices Sigma.7 with 7DoF of motion dedicated to the left and right hand of the surgeon.