Bioelectrical impedance analysis
Bioelectrical impedance analysis is a method for estimating body composition, in particular body fat and muscle mass, where a weak electric current flows through the body, and the voltage is measured in order to calculate impedance of the body. Most body water is stored in muscle. Therefore, if a person is more muscular, there is a high chance that the person will also have more body water, which leads to lower impedance. Since the advent of the first commercially available devices in the mid-1980s the method has become popular, owing to its ease of use and portability of the equipment. It is familiar in the consumer market as a simple instrument for estimating body fat. BIA actually determines the electrical impedance, or opposition to the flow of an electric current through body tissues, which can then be used to estimate total body water, which can be used to estimate fat-free body mass and, by difference with body weight, body fat.
Accuracy
Many of the early research studies showed that BIA was quite variable, and it was not regarded by many as providing an accurate measure of body composition. Nevertheless, it is the 4-compartment model – and not BIA – that is regarded as the reference method in body composition analysis.Although the instruments are straightforward to use, careful attention to the method of use should be given.
Simple devices to estimate body fat, often using BIA, are available to consumers as body fat meters. These instruments are generally regarded as being less accurate than those used clinically or in nutritional and medical practice. They tend to under-read body fat percentage by approximately 5 kg on average, despite showing a linear correlation with MRI-based measurements of 0.75 and 0.81 for females and males respectively.
Dehydration is a recognized factor affecting BIA measurements, as it causes an increase in the body's electrical resistance, so has been measured to cause a 5 kg underestimation of fat-free mass i.e. an overestimation of body fat.
Body-fat measurements are lower when measurements are taken shortly after consumption of a meal, causing a variation between highest and lowest readings of body fat percentage taken throughout the day of up to 4.2% of body fat.
Moderate exercise before BIA measurements lead to an overestimation of fat-free mass and an underestimation of body fat percentage due to reduced impedance. For example, moderate intensity exercise for 90–120 minutes before BIA measurements causes nearly a 12 kg overestimation of fat-free mass, i.e. body fat is significantly underestimated. Therefore, it is recommended not to perform BIA for several hours after moderate or high-intensity exercise.
BIA is considered reasonably accurate for measuring groups, of limited accuracy for tracking body composition in an individual over a period of time, but is not considered sufficiently accurate for recording of single measurements of individuals.
Consumer-grade devices for measuring BIA have not been found to be sufficiently accurate for single-measurement use and are better suited for use to measure changes in body composition over time for individuals. The two-electrode, e.g., foot-to-foot or hand-to-hand measurement, is generally found to be less accurate than 4-electrode methods. Multiple electrodes, typically eight, may be used located on the hands and feet, allowing measurement of the impedance of the individual body segments arms, legs and torso. The advantage of the multiple electrode devices is that body segments may be measured simultaneously without the need to relocate electrodes. Results for some impedance instruments tested found poor limits of agreement and in some cases systematic bias in estimation of visceral fat percentage, but good accuracy in the prediction of resting energy expenditure when compared with more accurate whole-body magnetic resonance imaging and dual-energy X-ray absorptiometry.
Impedance is frequency-sensitive; at low frequency the electric current flows preferentially through extracellular water only, while at high frequency the current can cross cell membranes and hence flows through total body water. In bioimpedance spectroscopy devices resistance at zero and high frequency can be estimated and, at least theoretically, should provide the optimal predictors of ECW and TBW and hence body fat-free mass respectively. In practice, the improvement in accuracy is marginal. The use of multiple frequencies or BIS in specific BIA devices has been shown to have high correlation with DXA when measuring body fat percentage. The correlation with DXA can be as high as 99% when measuring fat-free mass, if strict guidelines are adhered to. It is important to recognize that correlation is not a measure of accuracy or method agreement, BIA methods typically exhibit 2 standard deviation limits of agreement with reference methods of around ±10%.
Historical background
The electrical properties of tissues have been described since 1872. These properties were further described for a wider range of frequencies on a larger range of tissues, including those that were damaged or undergoing change after death.In 1962, Thomasset conducted the original studies using electrical impedance measurements as an index of total body water, using two subcutaneously inserted needles.
In 1969, Hoffer concluded that a whole-body impedance measurement could predict total body water. The equation showed a correlation coefficient of 0.92 with total body water. This equation, Hoffer proved, is known as the impedance index used in BIA.
In 1983, Nyober validated the use of whole body electrical impedance to assess body composition.
By the 1970s the foundations of BIA were established, including those that underpinned the relationships between the impedance and the body water content of the body. A variety of single-frequency BIA analyzers then became commercially available, such as RJL Systems and its first commercialized impedance meter.
In the 1980s, Lukaski, Segal, and other researchers discovered that the use of a single frequency in BIA assumed the human body to be a single cylinder, which created many technical limitations in BIA. The use of a single frequency was inaccurate for populations that did not have the standard body type. To improve the accuracy of BIA, researchers created empirical equations using empirical data to predict a user's body composition.
In 1986, Lukaski published empirical equations using the impedance index, body weight, and reactance.
In 1986, Kushner and Scholler published empirical equations using the impedance index, body weight, and gender.
However, empirical equations were only useful in predicting the average population's body composition and was inaccurate for medical purposes for populations with diseases. In 1992, Kushner proposed the use of multiple frequencies to increase the accuracy of BIA devices to measure the human body as 5 different cylinders instead of one. The use of multiple frequencies would also distinguish intracellular and extracellular water.
By the 1990s, the market included several multi-frequency analyzers and a couple of BIS devices. The use of BIA as a bedside method has increased because the equipment is portable and safe, the procedure is simple and noninvasive, and the results are reproducible and rapidly obtained. More recently, segmental BIA has been developed to overcome inconsistencies between resistance and the body mass of the trunk.
In 1996, an eight-polar stand-on BIA device, , that did not utilize empirical equations was created and was found to "offer accurate estimates of TBW and ECW in women without the need of population-specific formulas."
In 2018, AURA Devices brought the fitness tracker AURA Band with built-in BIA.
In 2020 BIA became available for Apple Watch users with the accessory AURA Strap with built-in sensors.
By the early 2020’s smartwatches like the Samsung Galaxy Watch 4 contained built in BIA’s.